Fully Automatic Open Cup Flash Point and Ignition Point Tester This instrument primarily determines the minimum temperature at which petroleum products and other flammable liquids, when mixed with air under specified conditions, flash ignite and burn upon encountering an ignition source. The open cup flash point tester is widely used in railways, aviation, power generation, petroleum industries, and research institutions to ensure the safe storage, transportation, and use of materials.
The open cup flash point tester heats according to the temperature curve specified in GB/T 3536—2008. It automatically ignites when the temperature approaches the preset flash point value. Upon detecting the flash point and fire point, the instrument automatically locks the results, displays and prints them, while simultaneously activating the fan to automatically cool the heater.
High-definition Android touchscreen with real-time temperature curve display and Chinese/English language switching.
Complies with applicable standards and offers broad application scope.
Automatic electric ignition: ignites automatically when temperature approaches preset flash point value; locks results upon flash point occurrence.
System self-check function monitors arm, sweep rod, ignition wire, printer, fan, etc.
Built-in thermal printer delivers quieter, faster, and clearer printing.
Display | 7-inch HD Android touchscreen, 1024*600 pixels |
Measurement Range | Ambient temperature ~ 400℃ |
Temperature Sensor | Class A PT100 (three-wire system, temperature compensated) |
Temperature Detection Accuracy | 0.1℃ |
Barometer Accuracy | ≤0.5kPa |
Repeatability | ≤4℃ |
Heating Rate | 14℃-17℃/min before expected flash point of 56℃; 5°C-6°C/min from 23°C before expected flash point |
Ignition Method | Electrical Ignition |
Flashback Detection Method | Flame Ionization Detection |
Ignition Point Function | Automatic Ignition Point Value Locking |
Cooling Method | Forced Air Cooling |
Temperature Control | Intelligent Temperature Control |
Display Content | Real-time Temperature Curve, Chinese/English language switching |
Data Storage | Unlimited, based on EMMC capacity |
Additional Features | Data export, cloud platform upload |
Printer | Thermal, supports Chinese characters, graphics, QR codes, unlimited lines |
Self-Diagnostic Function | Detects arm, scanning bar, ignition wire, printer, fan, etc. |
Power Supply | AC220V±10% 50Hz±5% |
Power Consumption | ≤600W |
Ambient Temperature | 10°C~35°C |
Ambient Humidity | <85% |
Dimensions & Weight | 43cm*28cm*35cm (L×W×H), 11.3kg |