Handheld Particle Counter

Handheld Particle Counter

  • Product Number:LD-SCL
  • Product Difference:A portable detection device specifically designed for rapid measurement of the quantity and size distribution of dust, particulate matter, and other microscopic particles in the air

Product Introduction

Handheld dust particle counters are portable detection devices specifically designed for rapid measurement of the quantity and size distribution of dust, particulate matter, and other microscopic particles in the air. These instruments are widely used across multiple industries, including but not limited to pharmaceutical manufacturing, semiconductor production, hospitals, cleanrooms, and air purifier performance verification.

Compliance Standards: JJF1190-2008, GB/T6167-2007

Cleanliness Standards: National Standard, ISO, 209E, New GMP (Dynamic/Static)

Standard Features: Audit trail functionality, supports multi-user and multi-level management.

Product Features

  • Handheld design for easy operation, delivering precise and reliable data ideal for field use.

  • Storage: Automatically records and saves sampling data with 4GB of built-in storage space.

  • Information Display: Real-time concentration bar graph, equivalent concentration calculation, real-time airflow value, and actual cleanliness grade.

  • Data storage capacity: 1–100,000 entries, queryable, supports USB data export.

  • Printer: Equipped with a dedicated Bluetooth printer for real-time data printing.

  • Network: Built-in Wi-Fi enables remote data transmission.

Product parameters

Light SourceSemiconductor laser, lifetime greater than 35,000 hours
Sampling Flow Rate① Standard Flow Rate: 2.83 L/min (0.1 CFM)
② Accuracy: ±5% (at 2.83 L/min)
Particle Counting① Particle Size Channels: 0.3μm, 0.5μm, 1.0μm, 2.5μm, 5.0μm, 10μm
②Maximum Count Capacity: 35,000 particles/liter (≥0.5μm particle size)
③Repeatability Relative Error: <10%
④Particle Size Distribution Error: ±30% for 0.5μm size band distribution
Display Method①Screen: Capacitive touch LCD
② Information Display: Real-time concentration bar graph, equivalent concentration calculation, real-time airflow value, actual cleanliness grade value
③ Additional Information: Time, date, measurement values, temperature/humidity, room number, sampling point, sampling count, battery level, status, etc.
④ Data storage capacity: 1–100,000 entries, queryable and exportable
Data processing① Storage: Automatically records and saves sampling data; built-in 4GB storage capacity
② Export: Supports USB flash drive data export and Bluetooth printer output of sampling data
③ Calculation: 95% UCL calculation, directly displays particle concentration (particles, particles/m³, particles/ft³)
BatteryLithium-ion battery, 3.7V, 10000mAh, battery life ≥8 hours
Operation & Connectivity①Interface: Dual USB high-speed communication ports for connecting USB drives and computers
②Network: Built-in Wi-Fi enables remote data transmission
③ Printer: Equipped with a dedicated Bluetooth printer for real-time data printing

Operating Environment

Temperature: 0°C to 40°C, Humidity: 10% to 75%, Sampling Gas: Must not contain corrosive gases such as acids or alkalis

Storage Environment

Temperature: 0°C to 45°C, Humidity: 0% to 90% RH

Alarm SettingsAlarm settings for cleanliness level alerts