The trace moisture analyzer employs the Karl Fischer titration principle and utilizes advanced microprocessor circuitry to determine trace moisture content in liquids and solids with varying properties. For solids insoluble in reagents, as well as substances that react chemically with reagents or readily contaminate electrodes, corresponding solid and liquid samplers can be used for indirect measurement. It is widely applied in industries such as power generation, petroleum, chemical engineering, pesticides, and pharmaceuticals.
High measurement accuracy, fast analysis speed, wide application range, stable and reliable performance, and simple operation.
The sample measurement process is automatically controlled by the instrument, completing stirring and measurement in approximately 60 seconds with direct display of results.
Fully sealed titration cell bottles prevent reagent contact with personnel and eliminate environmental humidity effects.
Utilizes advanced microprocessor circuitry to determine trace moisture content in liquids and solids with varying physical properties.
Display System | LED Five-Digit Decimal Display |
Titration Method | Electrolytic Titration Method (Coulometric Analysis) |
Electrolysis Current | Automatic Electrolysis Current Control (Maximum 400mA) |
Frequency | 50Hz ± 5Hz |
Measurement Range | 0.0001% (1ppm) to 100% |
Titration Speed | 0.6mg/min (Maximum) |
Sensitivity Threshold | 0.1 μg H₂O |
Operating Ambient Temperature | 5 to 40°C |
Power Supply | AC 220V ±10% |
Accuracy | 10 μg to 1 mg H₂O; ±3 μg >1 mgH₂O: RSD<0.3% (excluding sampling error) |