Halogen moisture analyzers are applicable to all industries requiring rapid moisture determination. They can measure any granular, powdered, flaked material, or non-volatile liquid, such as: grains, chemicals, pharmaceuticals, plastic raw materials, rubber, tobacco, tea, textiles, meat, foodstuffs, starch, feed, coal, foundry sand, sewage sludge, dairy products, and more.
Traditional moisture determination typically relies on oven drying methods. Testing a single sample can take two to three hours, or even three to four hours, requiring subsequent weighing on a balance and manual calculation to obtain the moisture content (water content). The inefficiency of oven-based moisture measurement fails to meet the demands of high-paced industrial production.
The Lucky Halogen Moisture Analyzer employs a drying heater—a high-quality annular halogen lamp—to rapidly and uniformly heat the sample, continuously evaporating its moisture. Throughout the measurement process, the instrument automatically displays real-time results: sample weight, moisture content, solids content, measurement time, heating temperature, and more.
The Halogen Moisture Analyzer employs the internationally recognized oven drying principle. Its measurement results demonstrate excellent consistency with oven-based moisture determination while achieving significantly higher efficiency. Most samples can be analyzed in just a few minutes, earning widespread user acclaim and positive reviews.
LCD touchscreen, compact and lightweight, requires no installation or calibration—ready to use right out of the box.
Even heating, stable performance, and accurate testing with short measurement times and high efficiency.
German-imported sensors deliver exceptional and precise testing results.
Fully automatic operation with audible alerts upon completion, requiring no supervision during measurement.
Requires no training with simple operation, eliminating cumbersome steps.
Extremely versatile, suitable for moisture determination across nearly all industries.
Unique draft shield design enhances weighing system stability.
Halogen lamp heating with pure stainless steel chamber housing for high-temperature resistance and easy cleaning.
Simultaneously converts sample moisture content and solids content.
Adjustable temperature/time settings with percentage display for moisture content and residue after drying.
Product Model | LD-LS1 | LD-LS2 | LD-LS3 | LD-LS4 |
Maximum capacity | 50g | 110g | 110g | 110g |
Reading accuracy | 0.005g | 0.005g | 0.002g | 0.001g |
Operating Temperature | 5°C---35°C | |||
Display | Touch LCD | |||
Calibration Method | External Calibration | |||
Scale pan dimensions | Φ90 | |||
External Dimensions (W × L × H) | 200*180*380 | |||
Product Weight | 5Kg | |||
Heating Method | Halogen Lamp | |||
Temperature Sensor | PT-100 | |||
Temperature Range | 40°C---199°C | |||
Temperature interval | 1°C | |||
Moisture Range | 0.00%---100% | |||
Moisture content readability | 0.0002 | 0.0002 | 0.0001 | 0.0001 |
Residual Drying Range | 100.00%---0.00% | |||
Dry residue remains readable | 0.0002 | 0.0002 | 0.0001 | 0.0001 |
ATRO | 100%---999% | |||
ATRO | 0%---999% | |||
Temperature Setting | 40°C---199°Cby1°Cstep | |||
Time Setting | 1---99min by 1s step | |||
Auto-stop | 0.1%---9.9%within 1min | |||
Storage History | 15 | |||
Heating Section | 220V+15% 50Hz | |||
115V+15% | ||||
400W | ||||
Connection Section | R232 Interface | |||
Balance Section | External power supply | |||
Output:9V | ||||
Lnput:220V+15%50Hz |