The Multi-Element Flame Photometer is a newly designed instrument. It features a compact size, user-friendly operation, and stable reliability. The entire series comes standard with an ion detection flame failure device, ensuring safety and dependability. It is widely used in the analysis and determination of agricultural fertilizers, soils, cement, ceramics, and other industries, as well as in the silicate industry and pathological research in medical and health fields. The flame photometer utilizes liquefied petroleum gas as its fuel.
Features direct concentration readout for five elements: K, Na, Li, Ca, and Ba, capable of simultaneously displaying up to three elements in any combination.
Equipped with an RS232 interface for connecting to printers and PC software.
Includes automatic correlation coefficient calculation, preset flame size selection, standard curve storage, and single-point calibration functions.
Equipped with flame failure protection and a moisture separator.
Features menu-driven keyboard operation.
Model | LD-22 | LD-30 | LD-31 | LD-32 | LD-40 | LD-41 | LD-50 | |
Display Range ug/ml | Potassium | 0-100 | 0-100 | 0-100 | 0-100 | 0-100 | 0-100 | 0-100 |
Sodium | 0-100 | 0-100 | 0-100 | 0-100 | 0-100 | 0-100 | 0-100 | |
Lithium | * | 0-100 | * | * | 0-100 | 0-100 | 0-100 | |
Calcium | * | * | 0-1000 | * | 0-1000 | * | 0-1000 | |
Barium | * | * | * | 0-3000 | * | 0-3000 | 0-3000 | |
Detection limit mmol/L | Potassium | ≤0.004 | ≤0.004 | ≤0.004 | ≤0.004 | ≤0.004 | ≤0.004 | ≤0.004 |
Sodium | ≤0.008 | ≤0.008 | ≤0.008 | ≤0.008 | ≤0.008 | ≤0.008 | ≤0.008 | |
Lithium | * | ≤0.015 | * | * | ≤0.015 | ≤0.015 | ≤0.015 | |
Calcium | * | * | ≤0.050 | * | ≤0.050 | * | ≤0.050 | |
Barium | * | * | * | ≤0.058 | * | ≤0.058 | ≤0.058 | |
Linear error mmol/L | Potassium | ≤0.005 | ≤0.005 | ≤0.005 | ≤0.005 | ≤0.005 | ≤0.005 | ≤0.005 |
Sodium | ≤0.03 | ≤0.03 | ≤0.03 | ≤0.03 | ≤0.03 | ≤0.03 | ≤0.03 | |
Lithium | * | ≤0.02 | * | * | ≤0.02 | ≤0.02 | ≤0.02 | |
Calcium | * | * | ≤0.07 | * | ≤0.07 | * | ≤0.07 | |
Barium | * | * | * | ≤0.07 | * | ≤0.07 | ≤0.07 | |
Stability | When continuously injecting standard solutions, the relative maximum change in the instrument reading within 15 seconds shall not exceed 3%. Measurements shall be taken once per minute, totaling six measurements. The relative maximum change in the instrument reading shall not exceed 15%. | |||||||
Response time | <8 seconds | |||||||
Minimum Suction Volume | <6mL∕min | |||||||
Repeatability | For the same standard solution, seven consecutive independent measurements must be ≤ 3% | |||||||
Test Element | Potassium Sodium | Potassium Sodium Lithium | Potassium Sodium Calcium | Potassium Sodium Barium | Potassium Sodium Lithium Calcium | Potassium Sodium Lithium Barium | Potassium Sodium Lithium Calcium Barium | |
Number of channels | 2 | 3 | 3 | 3 | 4 | 4 | 5 | |
Curve Calibration | Breakline method, linear approximation method, curve approximation method | |||||||
Printer | Optional | |||||||
Connect to the computer | Yes |