Multi-element Flame Photometry Analyzer

Multi-element Flame Photometry Analyzer

  • Product Number:LD-41
  • Product Difference:Analytical instruments for secondary fitting of K, Na, Li, and Ba

Product Introduction

The Multi-Element Flame Photometer is a newly designed instrument. It features a compact size, user-friendly operation, and stable reliability. The entire series comes standard with an ion detection flame failure device, ensuring safety and dependability. It is widely used in the analysis and determination of agricultural fertilizers, soils, cement, ceramics, and other industries, as well as in the silicate industry and pathological research in medical and health fields. The flame photometer utilizes liquefied petroleum gas as its fuel.

Product Features

  • Features direct concentration readout for five elements: K, Na, Li, Ca, and Ba, capable of simultaneously displaying up to three elements in any combination.

  • Equipped with an RS232 interface for connecting to printers and PC software.

  • Includes automatic correlation coefficient calculation, preset flame size selection, standard curve storage, and single-point calibration functions.

  • Equipped with flame failure protection and a moisture separator.

  • Features menu-driven keyboard operation.

Product parameters

ModelLD-22LD-30LD-31LD-32LD-40LD-41LD-50
Display Range ug/mlPotassium0-1000-1000-1000-1000-1000-1000-100

Sodium0-1000-1000-1000-1000-1000-1000-100

Lithium*0-100**0-1000-1000-100

Calcium**0-1000*0-1000*0-1000

Barium***0-3000*0-30000-3000
Detection limit mmol/LPotassium≤0.004≤0.004≤0.004≤0.004≤0.004≤0.004≤0.004

Sodium≤0.008≤0.008≤0.008≤0.008≤0.008≤0.008≤0.008

Lithium*≤0.015**≤0.015≤0.015≤0.015

Calcium**≤0.050*≤0.050*≤0.050

Barium***≤0.058*≤0.058≤0.058
Linear error mmol/LPotassium≤0.005≤0.005≤0.005≤0.005≤0.005≤0.005≤0.005

Sodium≤0.03≤0.03≤0.03≤0.03≤0.03≤0.03≤0.03

Lithium*≤0.02**≤0.02≤0.02≤0.02

Calcium**≤0.07*≤0.07*≤0.07

Barium***≤0.07*≤0.07≤0.07
StabilityWhen continuously injecting standard solutions, the relative maximum change in the instrument reading within 15 seconds shall not exceed 3%. Measurements shall be taken once per minute, totaling six measurements. The relative maximum change in the instrument reading shall not exceed 15%.
Response time<8 seconds
Minimum Suction Volume<6mL∕min
RepeatabilityFor the same standard solution, seven consecutive independent measurements must be ≤ 3%
Test ElementPotassium SodiumPotassium Sodium LithiumPotassium Sodium CalciumPotassium Sodium BariumPotassium Sodium Lithium CalciumPotassium Sodium Lithium BariumPotassium Sodium Lithium Calcium Barium
Number of channels2333445
Curve CalibrationBreakline method, linear approximation method, curve approximation method
PrinterOptional
Connect to the computerYes