Laser Particle Counter: Its confidence level measurement and verification standards comply with the conditions specified in the national standard GB/T6167-2007 and calibration specification JJF1190-2008, enabling automatic determination of cleanliness levels. A single sampling session can simultaneously measure particle counts across six size channels, displaying real-time particle counts and trend changes for each channel.
The laser dust particle detector meets the testing requirements for cleanrooms (zones) in precision manufacturing and testing across various industries, including: - Pharmaceutical sector (pharmaceutical plants, drug inspection institutes, hospitals, etc.) - Electronics industry (semiconductor factories, precision machinery production and processing, etc.) - Food hygiene sector (refined processing of dairy products, vacuum-sealed meat products, seasoning foods, agricultural products, etc.) - Filter manufacturers (quality and efficiency inspection of filters, etc.) - Optics and aerospace sectors.
Android 7.1 operating system for enhanced user interaction experience.
Built-in temperature/humidity sensor and ambient pressure sensor.
Core optical scattering sensor significantly eliminates stray light within the chamber, substantially improving signal-to-noise ratio. This enables more accurate and reliable measurements.
Configured with a monitoring platform, detection results can be wirelessly transmitted directly to the platform for long-term and short-term cleanliness analysis, aiding management.
Calculates particle counts convertible to m³ and ft³ based on 95% confidence level (UCL).
Automatically determines purification levels according to classification standards for user convenience.
Built-in next-generation high-speed thermal printer allows flexible printing of selected particle size channels.
Supports Wi-Fi transmission for wireless data upload; also supports USB flash drive data copying with plug-and-play functionality.
Dual-use AC/DC power with built-in rechargeable lithium battery, environmentally friendly and durable, supporting continuous sampling for over 6 hours.
Convenient and simple telescopic pan-tilt mount for easier installation and sampling operations.
Sampling Volume | 2.83 L/min (0.1 cfm) |
Six Particle Size Channels | 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 (μm) |
Self-cleaning Time | ≤10 min |
Particle Size Distribution Error | 0.5 μm - 5 μm ≤±30% |
Particle Concentration Indication Error | 0.5μm ≤±30% |
Measurement Repeatability Relative Deviation | ≤±10% |
Operating Environment | Ambient temperature 15~35℃, 10-85%RH |
Sampling Period | Freely configurable from 1 second to 99 minutes |
Sampling Delay | ≤99s |
Measurement Interval | ≤999s |
Sampling Mode | Single, ≤99 times (user-defined), continuous |
Light Source | Imported semiconductor laser diode (lifespan >30,000 hours) |
Display Method | 7-inch color touchscreen |
Data Storage Capacity | 10,000–100,000 entries, searchable and exportable |
Charging Time | 2 hours |
Design Operating Time | 8 hours |
Average Power | 7.5W |
Built-in Battery | 74Wh (16.8V/5Ah) |
Charger | 50W (16.8V/3A) |