The Volume and Surface Resistivity Tester is a high-precision, high-resistance measurement instrument designed for solid insulating materials, powders, and liquids. The instrument features an integrated shielded measurement chamber and incorporates a picoampere-level microcurrent acquisition module. With a maximum test voltage of 1,000 V and an upper resistance measurement limit of 10¹⁷ Ω, it comprehensively covers testing requirements ranging from conventional insulating materials to ultra-high-resistance materials.
It is primarily suitable for determining the volume and surface resistance and resistivity of solid insulating materials (such as plastics, rubber, laminated materials, films, resins, mica, ceramics, glass, insulating varnishes, and other insulating materials and components), powders, liquids, and various other insulating materials.
Wide Measurement Range: The measurement range extends from 10⁴ to 10¹⁷ Ω, meeting testing requirements for everything from conventional insulating materials to high-resistance materials.
Compliance with Standards: Fully compliant with national standards such as GB/T 31838.2 - 2019, GB/T 31838.3 - 2019, and the IEC 62631 international standard.
Integrated High-Voltage Source: Provides multi-step or continuously adjustable DC high-voltage output at 10 V, 50 V, 100 V, 250 V, 500 V, and 1000 V, with overcurrent and short-circuit protection.
High-Sensitivity Measurement: Integrates a picoampere (pA) sensitivity acquisition module to ensure testing accuracy for high-resistance materials.
Automation and Intelligence: Enables fully automatic control of voltage switching, range switching, and test timing; automatically calculates results; features simple operation; and reduces human error.
Data Output: Data is displayed on a large 7-inch screen, test data is printed in real time, and test records can be stored in multiple modes.
User-Friendly Interface: Equipped with a touchscreen that intuitively displays voltage, current, resistivity, and the test process.
| Item | Parameters |
| Resistance Measurement Range | 5×10⁴ Ω ~ 1×10¹⁷ Ω |
| Resistance Measurement Range and Accuracy | 5×10⁴ Ω ~ 1×10¹⁷ Ω |
| 5×10⁴ ~ 1×10⁸ Ω ±1% | |
| 1×10⁸ ~ 1×10¹⁰ Ω ±2% | |
| 1×10¹⁰ ~ 1×10¹² Ω ±5% | |
| 1×10¹² ~ 1×10¹⁷ Ω ±15–20% | |
| DC Test Voltage | 10/50/100/250/500/1000 V |
| DC Voltage Error | ±5% |
| Power Supply | AC 220 V/10 A / 50 Hz |
| Operating Power Consumption | Approx. 15 W |
| Operating Ambient Temperature | 0–40 °C |
| Operating Ambient Humidity | ≤70% |
| Control Method | 7-inch touchscreen operation |
| Dimensions | 408 mm × 322.5 mm × 287.5 mm |
| Weight | Approx. 9.6 kg |
| Data Output | Real-time printing of measurement results |
| Safety Protection | Overcurrent protection, Short-circuit protection |
| Shielded Structure | Integrated design of the measurement shielded chamber and the main unit |
| Data Storage | Three modes; 18 data entries per mode, with the ability to add or delete entries |