This instrument is designed for method development, precision quantification, and continuous measurement in research laboratories, pharmaceutical analysis, and third-party testing. It is suitable for analytical tasks with higher requirements for stray light, photometric error, and baseline stability. It employs a dual-beam C-T optical system that corrects for light source fluctuations via a reference beam and is equipped with a high-sensitivity PMT (photomultiplier tube), with a wavelength range of 190–900 nm. The maximum permissible error for transmittance is ±0.2%T, stray light is ≤0.03%T, and baseline flatness is ±0.0008 A. It features a 10.1-inch color capacitive touchscreen and supports spectral scanning, quantitative analysis, time-resolved scanning, and protein/DNA assays.
Suitable for UV-visible spectrophotometric analysis in research laboratories, pharmaceutical companies, third-party testing institutions, and quality control laboratories where high photometric accuracy, stray light control, and baseline stability are required. The dual-beam reference measurement configuration, combined with a PMT (photomultiplier tube), is well-suited for method development, precision quantification, continuous measurement, spectral scanning, and time-resolved scanning. It can be applied to tasks such as the content determination of drugs and pharmaceutical preparations, the analysis of fine chemicals and related solution systems, the testing of environmental and water quality indicators, as well as protein and DNA assays. The dual-beam configuration is particularly suitable for measurements requiring reference correction, long-term continuous data acquisition, or those that are sensitive to fluctuations in the light source. Specific detection capabilities should be determined based on the results confirmed by the relevant pharmacopoeia, standard methods, and laboratory procedures.
Dual-beam reference measurement system: Utilizes a dual-beam C-T optical configuration and a 1200 lines/mm holographic grating. By performing measurements and reference corrections using both the sample beam and the reference beam, it minimizes the impact of light source fluctuations on measurements and is suitable for continuous measurement and spectral scanning.
PMT Detection and Photometric Performance: Equipped with a high-sensitivity PMT (photomultiplier tube) featuring a built-in 200-step automatic gain control function, providing a wider dynamic range and higher signal-to-noise ratio. Maximum permissible error for transmittance is ±0.2%T (0–100%T); transmittance repeatability is ≤0.1%T; stray light is ≤0.03%T; baseline flatness is ±0.0008 A.
UV-Vis Band and 10.1-Inch Touchscreen: Automatic switching between deuterium and tungsten lamps, with a wavelength range of 190–900 nm. Equipped with a color capacitive touchscreen with a resolution of 1024×600, supporting both Chinese and English interfaces; the main unit can independently complete parameter settings, measurements, and result viewing.
Multimode Analysis and Biological Sample Measurement: Supports photometric measurement, multi-wavelength measurement, spectral scanning, quantitative measurement, time-resolved scanning, and protein/DNA quantification. Equipped with automatic wavelength calibration, baseline correction, peak detection, and standard curve fitting functions, it can perform routine photometric analysis and record absorbance changes over time.
Full Data Export and Report Generation: The standalone unit supports exporting all measurement data and test reports via a USB flash drive and generates A4-format PDF reports.
Spectral Bandwidth Configurable on Demand: The standard configuration features a 2 nm spectral bandwidth, with optional specifications of 1 nm, 1.8 nm, and 4 nm available; users can select the appropriate configuration based on the bandwidth required by the measurement method.
Multi-Pathlength and Quadruple Holder Expansion: Comes standard with a 10 mm standard cuvette holder; optional adjustable pathlength holders are available to accommodate pathlengths ranging from 10 to 100 mm. Additionally, an optional manual slide-out quadruple holder is available to facilitate sequential switching between multiple cuvettes, accommodating different pathlengths and sample loading arrangements.
Lamp Replacement Without Calibration and Maintenance Management: No optical path recalibration is required after replacing the light source. The system supports recording of light source usage time, making it easy to track usage and schedule maintenance. The main unit supports Wi-Fi connectivity for software updates.
| Parameter Name | Technical Specifications |
| Optical System | Dual-beam, C-T-type monochromator |
| Grating | 1200 lines/mm holographic grating |
| Detector | PMT (photomultiplier tube); |
| Light Source | Deuterium lamp + tungsten lamp, automatic switching |
| Wavelength Range | 190–900 nm |
| Spectral Bandwidth | 2 nm (standard); 1 nm, 1.8 nm, 4 nm (optional) |
| Maximum permissible wavelength error | ±0.1 nm (at 656.1 nm); ±0.3 nm (full wavelength range) |
| Wavelength repeatability | ≤0.1 nm |
| Wavelength resolution | 0.1 nm |
| Scanning speed | ≤2000 nm/min |
| Wavelength scanning speed | ≤9,000 nm/min |
| Absorbance range | -4.0 A to +4.0 A |
| Maximum allowable error in transmittance | ±0.2%T (0–100%T) |
| Transmittance repeatability | ≤0.1%T |
| Stray light | ≤0.03%T |
| 100%T noise (@500 nm) | ±0.0003 A; ±0.05%T; ≤0.0006 A; ≤0.1%T |
| Noise at 0%T | ±0.01%T; ≤0.02%T |
| Drift (@500 nm, 2-hour warm-up) | ≤0.1%T/h; ≤0.00045 A/h |
| Baseline Flatness (@ 200 nm – 850 nm) | ±0.0008 A |
| Display | 10.1-inch color capacitive touchscreen; Resolution 1024×600 |
| Interface Language | Chinese/English |
| Measurement Functions | Photometric measurement, multi-wavelength measurement, spectral scanning, quantitative measurement, time-resolved scanning, protein/DNA quantification |
| Operating Modes | Standalone operation; Supports PC-based control |
| Calibration and Data Processing | Automatic wavelength calibration, baseline correction, peak detection, calibration curve fitting |
| Data Interface | USB |
| Standalone Data Export | Export all measurement data and test reports via USB flash drive; PDF reports (A4 format) |
| Software Updates | Wi-Fi-enabled updates |
| Light Source Usage Log | Deuterium lamp, tungsten lamp |
| No-adjustment light source replacement | Supported |
| Standard cuvette holder | 10 mm standard cuvette holder |
| Optional cuvette holder | Adjustable pathlength holder (10–100 mm); Manual Push-Pull Quadruple Holder |
| Cuvette Specifications | 10 mm × 10 mm |
| Power Supply Voltage | AC 220 V ±22 V |
| Power Supply Frequency | 50 Hz ± 1 Hz |
| Power consumption | ≤110 W |
| Dimensions | 550 × 544 × 233 mm (W × D × H) |
| Weight | (31±1) kg |
| Operating Temperature | 15–35 °C |
| Operating Humidity | 30%–70% RH, non-condensing |
| Storage Temperature | -10–60 °C |